A quick-reference, four-chart guide for technicians preparing a wire-length investigation: what to issue, how to define the measured section, which readings to record and how to reset safely. For fuller curriculum practical resources, visit our Curriculum Experiments resources.
Scope: one uniform wire, one material and diameter, and a series of non-zero contact separations. This supports the wire-length element of AQA Physics required practical 3 and Trilogy required practical 15. Those activities also include resistor combinations, which are not covered by this guide alone. [1] [2]
P18 is a Better Equipped supporting-guide code, not “required practical 18”. It develops the wire work introduced in P03; P17 covers series/parallel measurements and P04 covers I–V characteristics. The seven-lead layout, 1.10 m wire cuts and calculator defaults are editorial preparation choices. Staff must approve the actual wire, current-limiting arrangement, minimum length and reading/cooling routine.
Define X–Y → measure paired V and I → switch off → compare resistance with length

One reusable wire station · complete series multiply records, not equipment
One complete series includes every planned length plus one return-to-reference V–I pair, on one record sheet. The reference recheck is not a new distinct length or an independent repeat. Extra sets receive the same series allocation. Expand each record for all points.
Seven detachable leads cover five connections around the series loop and two voltmeter branches. Wire is issued as 1.10 m cuts for an intended maximum 1.00 m span, subject to fixture fit. Lead, whole-wire-cut and paper spares round up separately. Reuse sound wire; additional investigations, cutting waste, PPE and spare components are separate. Settings and ticks are not saved.
Planning issue: 10 working stations + 1 extra = 11 equipped sets. Five distinct lengths + 1 reference recheck per series; 3 full series per set = 33 records and 198 planned V–I pairs. Leads: 77 working + 16 spares = 93. Wire: 11 working + 2 spare = 13 cuts of 1.10 m; 14.30 m before cutting waste. Paper: 33 records + 4 spares = 37 sheets. Reuse sound wire and apparatus after cooling/reset; counts do not set safe electrical limits.
| Item / issue specification | Per equipped set | Class provision (11 sets) |
|---|---|---|
| 1 | 11 | |
| 1 of each | 11 of each | |
| 1 | 11 | |
| 1 | 11 | |
| 1 complete kit | 11 kits | |
| 1 cut + class spares | 13 cuts (11 + 2 spare) 14.30 m total |
|
| 2 | 22 | |
| 7 + class spares | 93 leads (77 + 16 spare) |
|
| 1 connection kit | 11 kits | |
| 1 | 11 | |
| 3 / set | 37 sheets (33 + 4 spare) |
Labels, approved circuit card, suitable fixings/tape, cutters, staff check equipment, replacement fuses and assessed PPE. Micrometer work and different wire materials are optional additions.
The calculator does not select voltage, wire diameter, minimum separation, limiter resistance, power rating or cooling time. Confirm class service capacity and the actual connector layout.
Keep the same two endpoints for length and voltage · exclude the limiter from the measured span

Check wire identity, uniformity and damage. Mount it straight without tension and keep the ruler fixed. Decide one repeatable contact-position convention; do not use the jaw’s plastic housing as the endpoint.
With output off, check the complete loop, meter sockets and shared X/Y connections. Keep the high-input-resistance voltmeter across the active section; never use an Ω function on a powered circuit.
Check the shortest and longest planned spans for current, heating and readable voltage. Agree brief reading, cooling and a reference-return check. A steady display alone does not prove constant temperature.
| Part | Required arrangement | Avoid |
|---|---|---|
| Wire and tails | Same specimen, material and diameter; only active length changes. | Stretching, kinks, a new wire midway or a connected free tail. |
| Contact positions | Read the actual contact locations; start with the longest approved span. | Zero separation, broad inconsistent contacts or measuring the whole cut. |
| Meter connections | A in the feed; V on the same X–Y endpoints. Check actual V and I. | Using the supply dial as wire voltage or including the limiter in V. |
| Temperature routine | Low current, brief stable readings, output off and cooling; recheck reference. | Leaving power on between lengths or touching the wire to test heat. |
| Point / stage | xX / cm | xY / cm | L / m | V / V | I / A | R / Ω; notes |
|---|---|---|---|---|---|---|
| 1 · Reference | ____ | ____ | ____ | ____ | ____ | ____ |
| 2 · Length ____ | ____ | ____ | ____ | ____ | ____ | ____ |
| Further lengths | ____ | ____ | ____ | ____ | ____ | ____ |
| Return to reference | ____ | ____ | ____ | ____ | ____ | ____ |
Record wire ID/alloy/diameter, series ID, meter ranges and original display units. Convert position differences from cm to m and current from mA to A. Retain supply/limiter settings, cooling notes and all raw readings.
Check endpoints, temperature and units first · switch off before changing anything
| Problem | What to check | Technician action |
|---|---|---|
| No current or a zero display | Open circuit, incorrect socket/range, poor contact, failed fuse or an unsuitable display resolution. | Switch off and trace the circuit. Staff check the meter. Do not bypass a fuse or increase voltage blindly. |
| High current, heating or drifting values | Too-short span, incorrect limiter, a wire bypass or prolonged energising. | Switch off if safe, keep clear and alert staff. Inspect cold and re-pilot. Do not move live contacts or cool with water. |
| R changes at the same length | Wire heating, shifting/wide jaws, stretching, a kink or a different specimen. | Cool and reset using the same contact convention. Record the reference recheck and retain the original data. |
| Resistance is unexpectedly large or small | Wrong voltmeter span, mA treated as A, V/I inverted or near-zero readings. | Check the raw pair and units. Do not divide by zero; a zero current display is not evidence of zero resistance. |
| R–L graph misses the origin | Contact contributions, length offsets, temperature change, meter loading or limited resolution. | Use a justified fit and inspect the intercept. Do not invent a measured point at L = 0 or force the line through zero. |
| Repeats disagree or pattern is weak | Inconsistent contact placement, poorly resolved short spans, changed settings or cooling. | Compare the spread at each length. Re-pilot unsuitable ranges; label invalid data and explain exclusions. |
Public references checked: 21 October 2026. AQA scope and equations [1] [2] [3]; related IOP wire and V–I activities [4] [5] [6]; general HSE electrical guidance [7]. Full references and source/adaptation notes are below the printable area.
The station counts, contact convention and reference recheck are editorial choices requiring local review and a complete pilot. No external organisation has approved this guide.
Explore buying guides, practical resources and laboratory reference information.
This Technician Practical Essentials guide brings together apparatus issue, wire mounting, contact-to-contact length measurement, paired voltage/current readings and safe reset. It complements our Curriculum Experiments resources.
Source basis and adaptation: P17 supplies the established page style, preparation calculator behaviour and four-chart format. The wire method builds on P03’s wire-length route and the public teaching sources below. AQA establishes the required-practical scope; IOP describes related wire and V–I investigations. [1] [2] [4] [5] [6] The dedicated wire station, seven-lead allowance, five-length example and end-of-series reference check are editorial preparation choices, not a verbatim source procedure or a validated kit specification.
P18 is a focused supporting guide for resistance versus length. It addresses the wire-length element of AQA Physics required practical 3 and Trilogy required practical 15, but does not replace their resistor-combination work. P03 brings both elements together; P17 develops current and voltage measurements in series and parallel. P04 investigates I–V characteristics at a chosen component, which asks a different question. [1] [2] Check the specification used by your centre before mapping this guide to another awarding organisation.
The specimen can extend beyond both contacts, but only the section between X and Y is intentionally in the current path. Both voltmeter branches use those same endpoints. The position convention should identify the jaw–wire contact, not the plastic clip body. A broad jaw, changing grip or an uncertain contact point contributes to length uncertainty. Keep unused tails electrically free, clear of each other and safely retained.
Preparation example, not a prescribed range: a 1.10 m cut can allow a 1.00 m tested span plus short fixing tails, only when the actual mounting permits it. The example sequence of 100, 80, 60, 40 and 20 cm is not automatically safe or measurable with every alloy, diameter and supply. Record the actual contact separation rather than substituting the intended value. A reference contact need not coincide with the ruler’s zero.
L = |xY − xX|Use a consistent reference at each contact. Convert cm to m before plotting in SI units; the graph’s gradient then has units Ω/m. Never energise the circuit with coincident contacts.
The intended comparison changes length while keeping material, diameter and temperature sufficiently stable. Short spans can draw excessive current; IOP highlights heating and the value of low-current, short-duration readings. [4] [5] For this adaptation, staff set the supply/current limit and separate limiter using the actual kit, then trial the complete range. A suitable rheostat or fixed resistor assembly must have adequate current/power capacity; never assume its name or the supply’s low-voltage label establishes safety.
Readings should be brief but settled enough to resolve both meters. Switch off and use the approved cooling routine before moving the contact. Return to the initial reference length after the series to look for drift. Agreement is reassuring but does not prove the wire stayed at exactly constant temperature; disagreement can also come from contacts, meter behaviour or mechanical changes. Constantan’s name does not establish temperature independence.
There is no requirement in this guide to hold the supply dial, measured current or measured wire voltage artificially identical at every length. Any adjustment must follow the approved method and be recorded; each resistance comes from its own V–I pair. Keep the intended low-power regime comparable. The core method does not use water cooling, a bare copper short, or high-current heating.
Measurement-model note: with A upstream of the voltmeter branch, the ammeter reads wire current plus the small current drawn by V. If the voltmeter input resistance is RV, the idealised indicated V/I is the parallel combination of the tested path and RV. A sufficiently high RV makes this contribution small. This is a circuit-model explanation, not a correction to apply blindly to classroom data.
The two-contact arrangement is not a four-terminal precision resistance measurement. Contact contributions inside the sensing points and uncertain effective length can matter, especially when the wire voltage is very small. A non-zero graph intercept may be consistent with such effects, but cannot identify their cause by itself. Do not subtract an assumed contact resistance to make the results agree.
Use resistance vertically and measured length horizontally. Calculate R for every original pair, then compare valid repeats at the same length. Report the arithmetic mean of those calculated resistances and their spread when appropriate; do not silently replace it with mean voltage divided by mean current. Keep the return-to-reference check identifiable rather than counting it as another distinct length or an independent full repeat.
The uniform-wire model predicts proportionality between R and L when the other relevant properties remain fixed. [4] Keep an unconstrained, justified best fit and examine the intercept; a physically measured zero-length point would be a short circuit, not a useful data point. Different wire identities should not be pooled as repeat measurements of one specimen.
Ohm’s law is a different test. A resistance value is defined from V/I at a point; a linear I–V response requires an appropriate constant-condition investigation. Changing length while taking one V–I pair at each length does not establish that a single length is ohmic. [5]
Material or diameter comparisons require new labelled specimens, controlled lengths and fresh checks of safe current and measurement resolution. A department extending to resistivity may use R = ρL/A for a uniform wire; deriving ρ from an R–L gradient also requires an independently measured cross-sectional area. That is not what the base preparation calculator or core record determines. [4] An ohmmeter route is another separately approved method; never connect a resistance-function meter to this powered V–I circuit.
The example below stays fixed: 10 working stations + 1 extra set; five distinct lengths; three complete series per set; 20% spare leads, 10% spare wire cuts and 10% spare paper. The live calculator changes Chart 2, not these explanatory figures.
10 + 1 = 11 complete stations.
Two contacts per set = 22 contacts.
7 × 11 = 77 working leads.
20% × 77 = 15.4, rounded up to 16 spares.
Total = 93 leads. Check the actual connection style.
11 working cuts + ceiling(10% × 11) = 13 cuts.
13 × 1.10 m = 14.30 m before cutting waste.
Spare cuts do not add stations. Reuse sound wire after checks; a changed specimen needs its own identity.
11 × 3 = 33 complete-series records.
10% × 33 = 3.3, rounded up to 4 spare sheets.
Total = 37 record sheets.
Each sheet includes every length plus the return check.
5 length points + 1 reference recheck = 6 paired readings per series.
33 × 6 = 198 planned V–I pairs.
Extra fault investigations, invalid-reading replacements and separate student worksheets need their own allowance.
Review the actual supply, meters, leads, limiter, wire and contact fittings against current authorised school-science guidance and manufacturer instructions. Agree isolation, faults, cooling, sharp-end handling, supervision and assessed PPE. HSE provides general electrical-safety guidance, not approval of this station. [7]
No member-only CLEAPSS procedure or manufacturer instructions for the school’s actual equipment were reviewed for this draft. Public-source checking is separate from technical review and a physical pilot. Historical safety-test dates on teaching activities do not validate this adaptation.
Public sources checked 21 October 2026. Their scope is distinguished from the editorial station design.
No external source organisation has approved this guide. Complete local technical review and practical validation before classroom issue.
Have feedback or a useful technician tip? Please contact us.
Draft prepared: 21 October 2026 · Version: 1.0 · Technical review: pending confirmation before publication.
Build a compatible wire station around the actual resistance wire, contacts, protection and readable voltage/current measurements. Check ratings, mounting span and the locally approved method before ordering.
Links open in a new tab or window. These are catalogue searches, not verified product matches or current-stock guarantees. Confirm wire alloy/diameter, limiter current/power capacity, meter inputs and compatible contact connections. Optional diameter checks are not counted as one micrometer per station. A low-voltage label alone does not make a short wire safe.