Better Equipped Technician Practical Essentials
A quick-reference, six-page guide for technicians preparing GCSE anion tests: what to issue, how to keep reagent routes separate, what to record and how to clear down. For more detailed curriculum practical resources, visit our Curriculum Experiments resources.

Known single salts · fresh portions · observations before anion identification

Decision plan: normally check carbonate first, then sulfate and halides on fresh portions as appropriate. The supply checklist allows for all three routes; it does not require unnecessary tests or constitute a scheme for arbitrary mixtures.
Reusable wet-test sets · fresh portions per route · separate reagent budgets
Every set receives the same sample codes. One sample-round budgets for three fresh wet-test portions and the carbonate gas check, not a command to perform unnecessary tests. Four tubes per code cover carbonate, sulfate, halide and limewater; two further tubes are for local checks. Reuse clean apparatus between rounds.
Supply budgets per sample-round: sample stock 6 mL; HCl 3 mL; HNO3, AgNO3 and BaCl2 1 mL each; limewater 2 mL. Add the allowance, then round sample stock up to 1 mL per code per set and each reagent up to 5 mL per set before multiplying. These are supply budgets, not test doses or dilution recipes. Add common reference controls, reagent blanks, washing water and spare tubes separately. For different code allocations, calculate each class batch separately. Input limits are software bounds, not approved working quantities. Settings and ticks are not saved.
Interactive calculator not active. Fixed example: 10 stations + 1 extra set; two sample codes at each set, one analysis round and 25% handling allowance. Prepare 110 tubes and 88 mL of EACH sample. Reagents: 110 mL HCl; 55 mL EACH of HNO3, AgNO3, BaCl2 and limewater. Shared controls and blanks are additional. Adjust locally when scripting is unavailable.
Planning issue: 10 working stations + 1 extra = 11 sets; 2 sample codes/set × 1 round = 22 sample-rounds. Prepare 10 tubes/set: 110 total, reused between rounds. At 25% allowance, EACH sample needs 8 mL/set: 88 mL/code for the class. HCl: 10 mL/set, 110 mL class. HNO₃, AgNO₃, BaCl₂ and limewater: 5 mL/set EACH, 55 mL class EACH. Shared controls and blanks are additional.
| Item / issue specification | Per equipped set | Class provision |
|---|---|---|
| 10 | 110 | |
| 1 station kit | 11 kits | |
| 2 sample + 5 reagent | 22 sample + 55 reagent | |
| 1 assembly | 11 assemblies | |
| 1 preparation kit | 11 kits | |
| 8 mL / code | 88 mL / code | |
| 10 mL / set | 110 mL | |
| 5 mL / set | 55 mL | |
| 5 mL / set | 55 mL | |
| 5 mL / set | 55 mL | |
| 5 mL / set | 55 mL | |
| Per person | Count people separately |
Book known references, deionised-water reagent blanks, spare tubes, records and the staff identity key. The two local-check tubes are not a complete control set.
Book silver- and barium-waste collection, first rinsings, spill/first-aid arrangements, handwashing and supervision. No burners, ammonia or nitrate-test reagents are included.
Check stocks → pilot controls → separate acid routes → release the stations

Use school-selected soluble single salts with carbonate, sulfate, chloride, bromide or iodide; sodium salts are a practical example. Record concentrations and hazards. Begin with known references before coded unknowns; retain the staff key. No unlabelled powders or mixtures.
Use the actual stocks, tubes and gas-transfer parts. Confirm an observable carbonate gas test at the chosen scale, adequate acidification and distinguishable fresh halide precipitates. Check positive controls and reagent blanks; record permitted additions.
Give every sample and reagent a dedicated dropper. Keep HCl/BaCl₂ and HNO₃/AgNO₃ supplies visibly separate. Check clean glassware, white backgrounds, fresh limewater, stable supports and waste collection before issue.
| Ion | Reagent route | Positive observation / check |
|---|---|---|
| Carbonate · CO32− | Dilute HCl on a fresh portion; pass any gas through fresh clear limewater. | Cloudy / milky limewater supports CO₂. Fizzing alone is insufficient. |
| Sulfate · SO42− | Fresh portion: dilute HCl first, then BaCl2 solution. | White precipitate. No sulfuric acid; complete the piloted acidification first. |
| Chloride · Cl− | Fresh portion: dilute HNO3 first, then AgNO3 solution. | White precipitate. Keep chloride-containing reagents out of this route. |
| Bromide · Br− | Same nitric-acid / silver-nitrate route, on a fresh portion. | Cream precipitate; compare promptly with white and yellow references. |
| Iodide · I− | Same nitric-acid / silver-nitrate route, on a fresh portion. | Yellow precipitate. Do not identify from the starting solution colour. |
Use HCl → BaCl₂ for sulfate and HNO₃ → AgNO₃ for halides. Acidification removes carbonate interference within the chosen method. If effervescence continues, follow the piloted limit and ask staff before the next reagent.
Trial the complete small-volume route: dead space and gas escape can defeat the test. With a delivery line, keep the tip just below the limewater surface and the receiver open. Use the approved stopping sequence to prevent suck-back; never blow through the apparatus.
| Code / test | Reagents in order | Actual observation | Anion inference / next check |
|---|---|---|---|
| ____ / carbonate | |||
| ____ / sulfate | |||
| ____ / halide |
Check controls and contamination first · preserve evidence · segregate residues
| Problem | What to check | Technician action |
|---|---|---|
| White halide result in every tube | HCl or barium-chloride carry-over, tap-water residue, crossed droppers or dirty glassware. | Repeat the nitric-acid/silver-nitrate blank and known controls. Use fresh portions and clean tools; do not report every sample as chloride. |
| Unexpected sulfate precipitate | Wrong acid, incomplete carbonate removal, pre-existing cloudiness or an unsuitable sample. | Check the stock and acidification record. Repeat with the approved HCl route and controls. Never use sulfuric acid. |
| White, cream and yellow are unclear | Lighting, sample loading, dark backgrounds and ageing of the precipitate. | Compare fresh references side by side against white. Record ambiguous colour as inconclusive; do not add ammonia or heat the tube. |
| Little gas / no limewater change | Sample strength, scale, transfer losses, dead space, old limewater or a poor connection. | Use the known carbonate to check the whole route. Replace cloudy stock. No visible change can be inconclusive; do not enlarge the reaction arbitrarily. |
| Liquid draws back / line is blocked | Immersion depth, kinked tubing, poor supports or falling source pressure. | Stop and alert staff. Use the approved shutdown; never mouth-blow, force a joint or seal a reacting tube. Staff clear and test empty apparatus. |
| Results disagree across repeats | Reused portions, crossed codes, changed additions or different stocks. | Retain both records and check the controls. Repeat only as agreed. Do not select the expected result or infer an untested ion by elimination. |
Reference check: 12 October 2026. AQA curriculum [1]; RSC qualitative-test resources [2] [3] [4]. Full references and access limits are online. The sample bank, tube layout, aliquot examples and supply budgets are editorial adaptations requiring a local pilot and approval. No examination-board, RSC or CLEAPSS endorsement is claimed.


Explore buying guides, practical resources and laboratory reference information.
This Technician Practical Essentials guide brings together the issue quantities, three separate reagent routes, reference controls, observations, troubleshooting and clear-down for GCSE carbonate, sulfate and halide tests. It complements our Curriculum Experiments resources.
Source basis and adaptation: AQA supplies the curriculum tests and reference observations. [1] RSC supplies relevant teaching approaches and dilute reagent strengths. [2] [3] This guide uses a narrower anion-only bank, no ammonia and no heating. Its station layout, local-check tube count, small-scale portions and calculator budgets are editorial preparation choices, not a copied or laboratory-validated protocol.
C07 covers broader cation and anion identification. C11 gives more preparation detail for the anion routes only; C10 covers flame tests separately. AQA required practical 7 includes both positive and negative ions in unknown single ionic compounds. [1] C11 alone is not a complete substitute, and no universal Combined Science mapping is claimed.
A manageable bank uses school-known sodium carbonate, sodium sulfate, sodium chloride, sodium bromide and sodium iodide solutions. This is an editorial example, not a mandated stock list. The teacher records exact concentrations and confirms the selected tests by pilot. Coded samples come from this bank, not from unlabelled containers.
Show positive carbonate and sulfate controls and fresh chloride, bromide and iodide precipitates before asking for unknown identification. RSC’s negative-ion activity likewise starts with known observations and then moves to unknown solutions. [3] Use a reagent blank for each precipitation route: replace sample with deionised water, then follow the same acid and reagent sequence. A blank that forms a precipitate invalidates that run until contamination is resolved.
The default calculator represents two sample codes at every station; these can be selected known references or coded unknowns. It does not mean that every station receives the entire five-reference bank. Add common controls, their sample stock, test reagents and vessels separately. The two local-check tubes are spare working capacity, not a complete reference programme.
Hydrochloric acid introduces chloride, so it cannot acidify the halide portion without risking a false chloride result. Sulfuric acid introduces sulfate, so it is not suitable for the sulfate route. Use the selected HCl/BaCl₂ and HNO₃/AgNO₃ sequences, with separate aliquots and dedicated droppers. Keep the acid labels legible; a colour code alone is insufficient.
Carbonate can produce precipitates with the test reagents. The prior acid step removes it as carbon dioxide within the chosen method. The exact acid addition must be piloted against sample concentration; a generic “few drops” is not evidence that acidification is complete. Stop at the local limit and consult staff rather than improvising additional acid.
RSC’s small-scale carbonate example uses no more than 1 mL of limewater and transfers gas with a pipette. [3] C11 retains an up-to-1 mL limewater pilot example but leaves the gas-transfer arrangement to the school’s approved setup; a delivery-tube method is an adaptation, not the exact RSC method. Trial sample and acid quantities, gas escape, apparatus dead space and the first visible cloudiness using a known carbonate.
For a delivery tube, the receiving vessel stays open and the tip is only shallowly immersed. Stop and withdraw by the locally agreed procedure to prevent suck-back; never seal an outlet or mouth-blow through the apparatus. The supply budget of 2 mL limewater per sample-round is not an instruction to put 2 mL into that receiver. A failed transfer test can remain inconclusive even when effervescence is visible.
The example bank deliberately excludes hydrogencarbonates, sulfites, insoluble samples and unknown mixtures. A carbonate-style acid/gas response is not a universal identification scheme outside that bank. A halide colour supports the relevant anion under the tested conditions; it does not prove purity or determine the metal ion.
Provide a white background, consistent lighting and known fresh precipitates. Ask pupils to record colour words, turbidity, timing and reagent order. A photograph is supporting evidence, not a colour standard. Where colour discrimination is difficult, use teacher-supported comparison and keep the uncertainty explicit.
Use equations to explain the observed gas or precipitate after recording the evidence. They are not extra tests and do not justify adding new reagents. [1] [3]
CO₃²⁻(aq) + 2H⁺(aq) → CO₂(g) + H₂O(l)
CO₂(g) + Ca(OH)₂(aq) → CaCO₃(s) + H₂O(l)
The calcium carbonate precipitate accounts for the first cloudiness.
Ba²⁺(aq) + SO₄²⁻(aq) → BaSO₄(s)
Ag⁺(aq) + X⁻(aq) → AgX(s), where X is Cl, Br or I.
Record white, cream or yellow as appropriate; compare against the actual controls.
Sets = working stations + extra complete sets. A sample-round means provision for the planned analysis of one code once, with fresh portions for different test routes. More repeats increase solutions, not the number of reusable racks or gas-transfer assemblies. Extra sets include the same code allocation and run allowance. They do not guarantee sufficient supervision or lesson time.
Tubes per set = 4 × sample codes + 2. The four positions per code are carbonate reaction, limewater receiver, sulfate and halide; add two local-check positions. Racks must accommodate this many tubes, even when more than one rack is required. Sample and reagent stock bottles are separate from these tubes.
Sample stock: start with 6 mL per code per analysis round. Apply the percentage allowance, round up to 1 mL per code per set, then multiply by sets. Reagents: multiply the per-sample-round budget by codes and rounds, apply the allowance, round each reagent up to 5 mL per set, then multiply by sets. All are editorial supply allowances. They do not specify the volume added to a test, validate sample strength or replace an approved dilution procedure.
For 10 stations + 1 extra, two codes and one round with 25% allowance: 6 × 1 × 1.25 = 7.5 mL, rounded up to 8 mL of EACH sample per set. Eleven sets therefore need 88 mL of EACH sample, in separately labelled supplies. The combined volume for two different codes is 176 mL, not 88 mL total.
HCl: 3 × 2 × 1 × 1.25 = 7.5 mL, rounded to 10 mL/set; 110 mL for 11 sets. Each 1 mL-budget reagent rounds from 2.5 to 5 mL/set. Limewater rounds from 5 to 5 mL/set. Tubes: (4 × 2 + 2) × 11 = 110. Shared controls and blanks still need separate provision.
Check the concentration-specific supplier SDS, current school/employer guidance and the approved method for each selected solution. Assess splashes, glass breakage, soluble barium reagents, silver-containing residues, gas transfer, cleaning and first aid. Pupils use staff-prepared dilute solutions; stock preparation is a separate trained-staff procedure. No ammonia, concentrated-acid treatment, nitrate reduction, burner or heating step is included.
Keep silver- and barium-containing residues and contaminated first rinsings in their designated, compatible waste streams. Do not infer drain suitability from a clear solution or a negative test. Do not improvise bulk neutralisation. Complete the local pilot, safety assessment and technical review before issue; the page is a preparation reference, not a risk assessment.
Checked 12 October 2026. Public sources support the selected chemistry and curriculum. They do not endorse the C11 layout, sample bank or quantity calculator.
Consult current relevant CLEAPSS guidance through your school’s access. No member-only CLEAPSS guidance or product-specific SDS was reviewed for this draft; no external safety approval is claimed. The user’s supplied template governs the page design and four-chart structure, not the C11 chemical content.
Have feedback or a useful technician tip? Please contact us.
Draft prepared: October 2026 · Version: 1.0 · Technical review: pending confirmation before publication.
Build the issue around separate sample portions, clean tubes, dedicated droppers and clearly identified acid routes. Match equipment and dilute reagents to the approved school method; this is not a pre-approved kit.
Links open in a new tab or window. Catalogue searches do not guarantee a matching item or working concentration; current stock, packs and product SDS have not been verified here. Do not substitute a concentrated solution or solid directly into a pupil test. Check the apparatus list and local procedure before ordering.