A quick-reference, six-page guide for technicians preparing a GCSE titration: what to issue, how to prepare and rinse the apparatus, how to recognise the endpoint, what to record and how to clear down. For more detailed curriculum practical resources, visit our Curriculum Experiments resources.

Hydrochloric acid in the burette · sodium hydroxide in the flask · no heating

Tick-box issue checklist · equipment is reused between runs · extra sets are class-level
One station is one shared apparatus set; decide pupil grouping separately. Extra complete sets include equipment and a full run allowance for a demonstration, pilot or spare. Round the pilot acid titre up to a whole mL for planning; this is not an expected pupil result. Alkali is fixed at 25.0 mL per run. The percentage allows for conditioning, priming, burette remainder and handling losses; increase it if the pilot needs more. Each solution is rounded up to 10 mL per set, then multiplied by sets. This is a supply calculator, not a dilution recipe or a guarantee of concordance. Settings and ticks are not saved.
Interactive calculator not active. The fixed example is 10 stations plus 1 extra set, with 4 runs per set, 25 mL acid per run and 30% reagent allowance. Prepare 11 apparatus sets and 1,430 mL of each dilute solution. Adjust locally if JavaScript is unavailable.
Planning issue: 10 working stations + 1 extra complete set = 11 equipped sets. Allow 4 titrations per set, including the rough (44 runs). With 25 mL acid per run, 25 mL alkali per run and a 30% handling allowance, prepare 130 mL of each solution per set (1,430 mL of each for the class). Volumes are rounded up to 10 mL per set.
| Item / issue specification | Per equipped set | Class provision |
|---|---|---|
| 1 | 11 | |
| 1 assembly | 11 assemblies | |
| 1 | 11 | |
| 1 | 11 | |
| 1, reused | 11 | |
| 1 of each | 11 of each | |
| 2 | 22 | |
| 1 | 11 | |
| 1 of each | 11 of each | |
| 130 mL / set | 1,430 mL | |
| 130 mL / set | 1,430 mL | |
| 1 dropper supply | 11 supplies; fill locally | |
| Per person | Count people separately |
Labels and results sheets; teacher endpoint example; handwashing and spill/first-aid arrangements; identified waste containers. Provide any additional PPE required by the local assessment.
The nominal 0.10 mol/dm³ solutions, 30% allowance and four-run plan are editorial examples. Use the pilot, actual stock values and class experience to confirm quantities and time.
Check stock → pilot the endpoint → inspect the glassware → confirm the reading convention

Use approved dilute stocks, not unlabelled decants. Record the actual acid concentration, alkali batch and indicator formulation. Keep sodium hydroxide capped between uses; atmospheric carbon dioxide can alter it. A nominal label alone is not a verified standard. [7]
Use the class bottles, pipette and burette. Check a visible pink start and the first persistent colourless endpoint. Confirm a titre comfortably within the burette’s usable range, the repeat plan and the time needed. The 25 mL planning titre is not a pupil target.
Check for chips, a sticking tap, leaks and a blocked jet. Use a stable base and a suitable clamp; leave room for swirling the flask. Plan safe filling below eye level and accessible meniscus reading. Practise with water first where needed. [2]
| Apparatus | Final rinse | Why it matters |
|---|---|---|
| Burette | A little of the hydrochloric acid | Residual water would dilute the acid; also flush the tap and jet. |
| Volumetric pipette | A little of the sodium hydroxide | Residual water would dilute the measured alkali portion. |
| Conical flask | Deionised water only | A little water changes concentration, not alkali moles. Alkali rinsing would add unmeasured alkali. |
Burette upright; tap closed until use; no jet bubble or leak; filling funnel removed; initial reading recorded. Place the flask on the white tile with space to swirl and the jet clear of the liquid. Refill only between recorded runs. [3]
Same alkali aliquot, indicator formulation and drop count; same first persistent loss of pink. Rinse the flask between runs and record a fresh initial reading. Do not back-titrate an overshoot with extra alkali. [4]
| Run | Initial reading | Final reading | Titre: final − initial | Used in mean / reason |
|---|---|---|---|---|
| Rough | No — rough | |||
| Accurate 1 | ||||
| Accurate 2 | ||||
| Accurate 3 |
Check technique before changing stock · preserve raw readings · keep waste streams identified
| Problem | What to check | Technician action |
|---|---|---|
| Bubble in jet / leaking tap | Air below the tap, falling level with the tap closed, drips or a cracked jet. | Stop and make the apparatus safe. Prime or replace it, then take a fresh initial reading. Repeat the affected run; do not “correct” the titre by guessing. |
| No pink colour at the start | Correct alkali, indicator identity and condition, contamination and the measured aliquot. | Compare with the pilot. Resolve the cause before titrating; do not keep adding indicator or substitute a different colour rule without agreement. |
| Endpoint passed too quickly | Large additions, poor swirling or failing to slow down near the rough titre. | Keep and label the affected result. Restart with fresh alkali; approach dropwise. Do not add extra alkali to restore the pink colour. |
| Accurate titres disagree | Meniscus reading, last-drop control, consistent pipetting and indicator dose. | Rehearse the technique and obtain additional runs if approved. Use the agreed concordance rule, not the pair nearest an expected answer. |
| Titres drift between runs | Unconditioned burette/pipette, crossed beakers, open alkali bottles or dirty flask. | Review rinsing and stock labels. Use fresh, checked working portions; cap stocks. Do not return leftover reagent to the stock bottle. |
| Titre differs from the pilot | Different stock strengths, wrong pipette volume, transposed readings or an exhausted burette scale. | Retain the data and ask the teacher to check the setup and stock. Never change readings or concentrations to force the expected titre. |
Sources checked 9 October 2026: AQA practical assessment [1]; RSC titration, technique and interpretation resources [2] [3] [4] [5] [6]; RSC background on NaOH and air [7]. Full references and adaptation notes are online.
The stock concentrations, supply allowances and concordance example are editorial planning choices. Pilot and approve locally. No examination-board, RSC or CLEAPSS approval is claimed; member-only CLEAPSS guidance was not consulted.
Explore buying guides, practical resources and laboratory reference information.
This Technician Practical Essentials guide brings together apparatus issue, reagent planning, conditioning and filling, endpoint checks, results recording, troubleshooting and clear-down for a GCSE acid–alkali titration. It complements our Curriculum Experiments resources.
Source basis and adaptation: AQA supplies the curriculum scope; published RSC resources support the titration techniques. The exact stock example, supply calculator and concordance example are editorial preparation choices, not a verbatim published protocol or a laboratory-validated method. Use a school-approved, piloted procedure. [1] [2] [3]
AQA GCSE Chemistry required practical 2 distinguishes reacting-volume work from Higher Tier concentration calculations in mol/dm³ and g/dm³. It is Chemistry-only. Other awarding bodies and courses need separate mapping; this is not an additional examination-board requirement. [1]
C02 uses HCl in the burette, NaOH in the flask, a 25.0 mL aliquot, nominal 0.10 mol/dm³ stocks and phenolphthalein. These are editorial choices to pilot. RSC’s salt-preparation activity uses different nominal concentrations and can continue to salt isolation; C02 stops after titration and calculations. [2]
For quantitative work, identify which solution is the known reference and document its supplied, prepared or checked concentration using the school’s analytical procedure. Here that is hydrochloric acid. A nominal 0.10 mol/dm³ label is not by itself evidence of an exact concentration. Keep safety identification on pupil samples even when a numerical concentration is withheld for the exercise.
With acid added to alkali, the selected phenolphthalein changes from pink to colourless. Use the minimum consistent drop count that gives a clear observation. Its formulation may introduce additional hazards; check the actual SDS. Do not substitute universal indicator. An alternative indicator needs its own agreed colour criterion. [2] [3]
Water remaining in the burette or pipette dilutes the solution being measured, so condition these with their assigned reagent. A little deionised water in the receiving flask does not add or remove alkali moles; it can also wash splashes back into the reaction mixture. Keep that distinction clear. [3] [5]
Use the markings and instructions for the actual pipette, including any drain and wait time. The usual non-blow-out volumetric transfer pipette is not emptied by blowing its final residue into the flask. Check the burette scale and reading convention separately; displayed decimal places do not establish apparatus accuracy.
Suppose the verified hydrochloric acid concentration is 0.100 mol/dm³, the pipetted alkali volume is 25.0 mL and the accepted acid titres are 24.60 mL and 24.70 mL. These meet the illustrative “within 0.10 mL” rule. Their mean is 24.65 mL; no rough titre is included.
n(HCl) = 0.100 × (24.65 ÷ 1,000) = 0.002465 mol. The equation gives equal moles of NaOH. Therefore c(NaOH) = 0.002465 ÷ 0.0250 = 0.0986 mol/dm³. The conversion to dm³ is essential when using n = cV.
Using Mᵣ(NaOH) = 40.0, mass concentration = 0.0986 × 40.0 = 3.944 g/dm³, or 3.94 g/dm³ to three significant figures. Keep extra digits during working, then round the final answer appropriately.
The calculation is an original numerical example using the 1 : 1 HCl/NaOH equation. Sulfuric acid would require a different mole ratio; do not reuse this shortcut unchanged. Concentration calculations belong to the Higher Tier element of the AQA practical. [1] [5]
Equipped sets = working stations + extra complete sets. Each set receives one reusable titration station and the selected number of runs, including the rough. Two runs only allow a rough plus one accurate result; select more runs to plan for a concordant pair. Extra runs are a supply provision, not a requirement to include every result in the mean.
Acid per set = runs × pilot-based acid planning volume × (1 + allowance ÷ 100), rounded up to the next 10 mL. Alkali per set uses the same formula with 25.0 mL per run. Class totals multiply each rounded per-set volume by the number of sets. The default is 11 sets × 130 mL = 1,430 mL of each solution. Equipment is not multiplied by the number of runs.
The handling allowance must cover conditioning, priming, unused liquid below the last reading and local losses. Repeatedly filling to zero and discarding the remainder can use more than this estimate. Pilot the actual issue routine; adjust allowance and stock separately. Changing the selected number of runs or sets does not change the approved concentration, pipette volume or risk assessment. Eyewear is counted per person and indicator supply is filled locally, not calculated from a universal drop volume.
Checked 9 October 2026. These public sources support the chemistry, curriculum and technique. They do not validate the calculator’s editorial stock allowances or replace a local risk assessment.
Consult current relevant CLEAPSS guidance through the school’s access, together with the SDS for the actual acid, alkali and indicator. Member-only advice has not been consulted here. No external endorsement is implied.
Have feedback or a useful technician tip? Please contact us.
Draft prepared: October 2026 · Version: 1.0 · Technical review: pending confirmation before publication.
Build the issue around a reliable burette, a matching volumetric pipette and filler, a stable stand and clean receiving glassware. Use the checklist above to confirm quantities, and choose chemicals and indicator formulation to the school’s approved method. No heating equipment is needed for this titration.
Check delivery markings, filler compatibility, clamp fit and readable burette graduations before selecting equipment. These are individual catalogue options, not a pre-approved complete kit.
Links open in a new tab or window. Direct product links were identified in public search on 9 October 2026; full product-page access was restricted during preparation. Check current specifications, packs and availability before ordering. Search links do not guarantee a matching concentration, formulation or item.